CALCulate Subsystem FSE
1065.6016.12
6.40
E-16
:CALCulate<1|2>:MARKer<1 to 4>:COUNt ON | OFF
This command switches on or off the frequency counter at the marker position.
Example:
":CALC:MARK:COUN ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
A
:CALCulate<1|2>:MARKer<1 to 4>:COUNt:RESolution 0.1 | 1 | 10 | 100 | 1000 | 10000 Hz
This command specifies the resolution of the frequency counter.
Example:
":CALC:MARK:COUN:RES 1kHz"
Features:
*RST value:
1kHz
SCPI:
device-specific
Mode:
A
The numeric suffix in MARKer<1 to 4> is not significant.
:CALCulate<1|2>:MARKer<1 to 4>:COUNt:FREQuency?
This command queries the result of the frequency counter.
Example:
":CALC:MARK:COUN:FREQ?"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, BTS, MS
This command is only a query and thus has no *RST value.
:CALCulate<1|2>:MARKer<1 to 4>:COUPled[:STATe] ON | OFF
This command switches the coupling of markers on or off.
Example:
":CALC:MARK:COUP ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Modes:
VA
The numeric suffix in MARKer<1 to 4> is not significant.
:CALCulate<1|2>:MARKer<1 to 4>:LOEXclude ON | OFF
This command switches the local oscillator suppression on or off.
Example:
":CALC:MARK:LOEX ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
A-F
The numeric suffixes 1|2 and 1 to 4 are not significant.
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......