
CONFigure-Subsystem FSE
1065.6016.12
6.84
E-16
:CONFigure[:MS]:SWEeptime STANdard | AUTO
This command selects the sweep-time computing mode for the spurious measurement:
Example:
":CONF:SWE AUTO"
Feature:
*RST value:
STANdard
SCPI:
device-specific
Mode:
MS
STANdard
The computation of the sweep time is based on a worst-case estimation
AUTO
The sweep time is reduced by a factor of 8 (assuming all slots are on).
:CONFigure[:MS]:MTYPe GMSK | EDGE
This command selects the standard (GMSK or EDGE).
Example:
":CONF:MTYP EDGE"
Features:
*RST value:
GMSK
SCPI:
device-specific
Mode:
MS
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......