FSE DISPlay Subsystem
1065.6016.12
6.95
E-16
:DISPlay[:WINDow<1|2>]:TIME ON | OFF
This command switches on or off the screen display of date and time.
Example:
":DISP:TIME ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Modes:
A, VA, BTS, MS
The numeric suffix in WINDow<1|2> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:X[:SCALe]:RVALue <numeric_value>
This command defines the reference value for the X-axis of the measuring diagram.
Example:
":DISP:TRAC:X:RVAL 20SYM"
Features:
*RST value:
-
SCPI:
device-specific
Mode:
VA-D
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:X[:SCALe]:ZOOM ON | OFF
This command displays the zoomed current frequency range in the opposite window of the split
screen representation.
Example:
":DISP:TRAC:X:ZOOM ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Mode:
A-F
The numeric suffix in TRACe<1 to 4> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:X[:SCALe]:ZOOM[:FREQuency]:STARt
<numeric_value>
This command defines the start frequency of the zoomed display range. The value must lie between
start and stop frequency of the original range.
Example:
":DISP:TRAC:X:ZOOM:STAR 100MHZ"
Features:
*RST value:
-- (depending on the current frequency setting)
SCPI:
device-specific
Mode:
A-F
The numeric suffix in TRACe<1 to 4> is not significant.
:DISPlay[:WINDow<1|2>]:TRACe<1 to 4>:X[:SCALe]:ZOOM[:FREQuency]:STOP <numeric_value>
This command defines the stop frequency of the zoomed display range. The value must lie between
start and stop frequency of the original range.
Example:
":DISP:TRAC:X:ZOOM:STOP 200MHZ"
Features:
*RST value:
-- (depending on the current frequency setting)
SCPI:
device-specific
Mode:
A-F
The numeric suffix in TRACe<1 to 4> is not significant.
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......