FSE SENSe Subsystem
1065.6016.12
6.191
E-16
:[SENSe<1|2>:]FILTer:LPASs:FREQuency <numeric_value>
This command defines the frequency limit of the low-pass filter in the NF-branch for analog
demodulation.
Parameter:
<numeric_value> ::=
3 kHz | 15 kHz for REAL TIME ON
5 PCT | 10 PCT | 25 PCT for REAL TIME OFF
Example:
":FILT:LPAS:FREQ 3KHZ
" for REAL TIME ON
":FILT:LPAS:FREQ 25PCT
" for REAL TIME OFF
Features:
*RST value:
- (STATe = OFF)
SCPI:
conforming
Mode:
VA-A
:[SENSe<1|2>:]FILTer:CCITt[:STATe] ON | OFF
This command activates the weighting filter according to CCITT-recommendation in the AF-branch
for analog demodulation.
Example:
":FILT:CCIT ON"
Features:
*RST value:
OFF
SCPI:
conforming
Mode:
VA-A
This command is valid only in vector analysis mode with REAL TIME OFF.
:[SENSe<1|2>:]FILTer:CMESsage[:STATe] ON | OFF
This command activates the C-message weighting filter according to CCITT-recommendation in the
AF-branch for analog demodulation.
Example:
":FILT:CMES ON"
Features:
*RST value:
OFF
SCPI:
conforming
Mode:
VA-A
This command is valid only in vector analysis mode with REAL TIME OFF.
:[SENSe<1|2>:]FILTer:DEMPhasis[:STATe] ON | OFF
This command activates the selected de-emphasis for analog demodulation.
Example:
":FILT:DEMP ON"
Features:
*RST value:
OFF
SCPI:
conforming
Mode:
VA-A
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......