INPut Subsystem FSE
1065.6016.12
6.128
E-16
:INPut<1|2>:ATTenuation:STEPsize 1dB | 10dB
This command defines the attenuator stepsize.
Example:
":INP:ATT:STEP 1dB"
Features:
*RST value:
10dB
SCPI:
device-specific
Mode:
A, VA, BTS, MS
This command is only available in conjunction with option FSE-B13, 1-dB attenuator.
:INPut<1|2>:UPORt<1|2>[:VALue]?
This command queries the control lines of the user ports.
Example:
":INP:UPOR2?"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, VA, BTS, MS
This command is a query command which is why it is not assigned an *RST value.
:INPut<1|2>:UPORt<1|2>:STATe ON | OFF
This command toggles the control lines of the user ports between INPut and OUTPut.
Example:
":INP:UPOR2:STAT ON"
Features:
*RST value:
ON
SCPI:
device-specific
Modes:
A, VA, BTS, MS
With ON, the user port is switched to INPut, with OFF to OUTPut.
:INPut<1|2>:IMPedance 50 | 75
This command sets the nominal input impedance of the instrument.
Example:
":INP:IMP 75"
Features:
*RST value:
50
SCPI:
conforming
Modes:
A, VA, BTS, MS
Switching the input impedance to 75
Ω
includes the matching elements RAM or RAZ selected by the
command
INPut:IMPedance:CORRection
.
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......