FSE STATus Subsystem
1065.6016.12
6.225
E-16
:STATus:QUEStionable:TRANsducer:PTRansition 0 to 65535
This command sets the edge detectors of all bits of the :STATus:QUEStionable:TRANsducer
register from 0 to 1 for the transitions of the CONDition bit.
Example:
":STAT:QUES:TRAN:PTR 65535"
Features:
*RST value:
–
SCPI:
device-specific
Modes:
A, VA, BTS, MS
:STATus:QUEStionable:TRANsducer:NTRansition 0 to 65535
This command sets the edge detectors of all bits of the :STATus:QUEStionable:TRANsducer
register from 1 to 0 for the transitions of the CONDition bit.
Example:
":STAT:QUES:TRAN:NTR 65535"
Features:
*RST value:
–
SCPI:
device-specific
Modes:
A, VA, BTS, MS
:STATus:QUEue[:NEXT]?
This command queries the earliest entry to the error queue, thus deleting it.
Example:
":STAT:QUE?"
Features:
*RST value:
–
SCPI:
conforming
Modes:
A, VA, BTS, MS
Positive error numbers indicate device-specific errors, negative error numbers are error messages
defined by SCPI (cf. Chapter 9). If the error queue is empty, the error number 0, "no error", is
returned. This command is identical with the command
SYSTem:ERRor
.