CALCulate Subsystem FSE
1065.6016.12
6.62
E-16
CALCulate:X and CALCulate:UNIT Subsystem
The CALCulate:X and CALCulate:Unit subsystems define the units for vector signal analyzer mode and
power measurements.
COMMAND
PARAMETERS
UNIT
COMMENT
CALCulate<1|2>
:X
:UNIT
:TIME
:UNIT
:ANGLe
:POWer
S | SYM
DEG | RAD
DBM | V | W | DB |
PCT | UNITLESS |
DBPW | WATT |
DBUV | DBMV | VOLT |
DBPT | DBUA | AMPere
DBUV_MHZ | DBMV_MHZ |
DBUA_MHZ | DBUV_M | DBUA_M |
DBUV_MMHZ | DBUA_MMHZ
Vector Signal Analysis
Vector Signal Analysis
:CALCulate<1|2>:X:UNIT:TIME S | SYM
This command selects seconds or symbols as an x-axis unit.
Example:
":CALC:X:UNIT:TIME S"
Features:
*RST value:
S
SCPI:
device-specific
Mode:
VA-D
:CALCulate<1|2>: UNIT:ANGLe DEG | RAD
This command selects the unit for angular measurement.
Example:
":CALC:UNIT:ANGL DEG"
Features:
*RST- value:
RAD
SCPI:
device-specific
Mode:
VA-D
:CALCulate<1|2>: UNIT:POWer
DBM | V | W | DB | PCT | DBPT | UNITLESS | DBPW | WATT |
DBUV |DBMV | VOLT | DBUA | AMPere | DBUV_MHZ |DBMV_MHZ|
DBUA_MHZ | DBUV_M | DBUA_M |DBUV_MMHZ | DBUA_MMHZ
This command selects the unit for power.
Example:
":CALC:UNIT:POW DBM"
Features:
*RST value:
_
SCPI:
device-specific
Modes:
A, VA
The units DEG, RAD, S, and HZ are available only in the vector analyzer mode.
DBUV_MHZ and DBUA_MHZ denote the units DBUV/MHZ or DBUA/MHZ.
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......