ABORt / CALCulate Subsystem FSE
1065.6016.12
6.8
E-16
ABORt Subsystem
The ABORt subsystem contains the commands for aborting triggered actions. An action can be
triggered again immediately after being aborted. All commands trigger events which is why they are not
assigned any *RST value.
COMMAND
PARAMETERS
UNIT
COMMENT
ABORt
--
--
no query
:ABORt
This command aborts a current measurement and resets the trigger system.
Example:
"
ABOR;INIT:IMM"
Features:
*RST value:
0
SCPI:
conforming
Modes:
A, VA, BTS, MS
CALCulate Subsystem
The CALCulate subsystem contains commands for converting instrument data, transforming and
carrying out corrections. These functions are carried out subsequent to data acquistion, i.e., following
the SENSe subsystem.
In the split-screen representation, a distinction is made between CALCulate1 and CALCulate2:
CALCulate1 =
^ screen A;
CALCulate2 =
^ screen B
For setting REAL/IMAG PART in Vector Analyzer mode a distinction is also made between CALCulate3
and CALCulate4 in the split-screen representation:
CALCulate3 =
^ screen C;
CALCulate4 =
^ screen D