FSE FETCh-Subsystem
1065.6016.12
6.107
E-16
:FETCh:BURSt:MACCuracy:RMS:STATus?
This command reads out the status of the RMS measurement of the modulation accuracy taken over
the selected number of bursts.
0: failed, 1: passed
Example:
":FETC:BURS:MACC:RMS:STAT?"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see
CONFigure:BURSt:MACCuracy
).
:FETCh:BURSt:MACCuracy:RMS:AVERage?
This command reads out the average of the RMS measurement of the modulation accuracy taken
over the selected number of bursts.
Example:
":FETC:BURS:MACC:RMS:AVER?"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see
CONFigure:BURSt:MACCuracy
).
:FETCh:BURSt:MACCuracy:RMS:MAXimum?
This command reads out the maximum of the RMS measurement of the modulation accuracy taken
over the selected number of bursts.
Example:
":FETC:BURS:MACC:RMS:MAX?"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see
CONFigure:BURSt:MACCuracy
).
:FETCh:BURSt:MACCuracy:PEAK:STATus?
This command reads out the status of the PEAK measurement of the modulation accuracy taken
over the selected number of bursts.
0: failed, 1: passed
Example:
":FETC:BURS:MACC:PEAK:STAT?"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
BTS, MS
If no measurement has been performed yet, a query error results.
This command is a query only and therefore has no *RST value assigned. It is available only if
measurement of the modulation accuracy is selected (see
CONFigure:BURSt:MACCuracy
).
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......