
FSE SENSe Subsystem
1065.6016.12
6.207
E-16
SENSe:SWEep Subsystem
This subsystem controls the sweep parameters.
COMMAND
PARAMETERS
UNIT
COMMENT
[SENSe<1|2>]
:SWEep
:TIME
:AUTO
:COUNt
:EGATe
:LEVel
:TYPE
:POLarity
:HOLDoff
:LENGth
:SOURce
:GAP
:PRETrigger
:TRGTogap
:LENGth
:SPACing
<numeric_value>
<Boolean>
<numeric_value>
<Boolean>
<numeric_value>
LEVel | EDGE
POSitive|NEGative
<numeric_value>
<numeric_value>
EXTernal | RFPower
<Boolean>
<numeric_value>
<numeric_value>
<numeric_value>
LINear | LOGarithmic
S
--
--
--
V
--
--
S
S
--
--
S
S
S
:[SENSe<1|2>:]SWEep:TIME <numeric_value>
This command defines the duration of the sweep.
The allowed value range depends on the selected domain and, for frequency domain, on the set
resolution bandwith (see Chapter 4, Softkey SWEEP TIME MANUAL).
Example:
":SWE:TIME 10s"
Features:
*RST value
- (AUTO is set to ON)
SCPI:
conforming
Modes:
A, VA-A
If
SWEep:TIME
is directly programmed, automatic coupling is switched off.
:[SENSe<1|2>:]SWEep:TIME:AUTO ON | OFF
This command switches the automatic coupling of the sweep time for the frequency span or
bandwidth settings.
Example:
":SWE:TIME:AUTO ON"
Features:
*RST value:
ON
SCPI:
conforming
Mode:
A
If
SWEep:TIME
is directly programmed, automatic coupling is switched off.
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......