![Rohde & Schwarz 1065.6000.20 Скачать руководство пользователя страница 121](http://html1.mh-extra.com/html/rohde-and-schwarz/1065-6000-20/1065-6000-20_operating-manual_1477758121.webp)
CALibration Subsystem FSE
1065.6016.12
6.64
E-16
:CALibration:LDETector?
This command performs a calibration of the log module’s characteristic and of the detectors. A "0" is
returned if the calibration was successful.
Example:
":CAL:LDET?"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, VA, BTS, MS
:CALibration:LOSuppression?
This command performs a calibration of the local oscillator suppression. . A "0" is returned if the
calibration was successful.
Example:
":CAL:LOS?"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, VA, BTS, MS
This command is only valid by model ‘30 instruments or by instruments which have been retrofitted.
:CALibration:PPEak?
This command performs a calibration of the tracking YIG filter (preselector peaking). A "0" is
returned if the calibration was successful.
Example:
":CAL:PPE?"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A
This command is only valid by the models FSEM and FSEK.
:CALibration:SHORt?
This command performs a short calibration. A "0" is returned if the calibration was successful.
Example:
":CAL:SHOR?"
Features:
*RST value:
-
SCPI:
device-specific
Modes:
A, VA, BTS, MS
:CALibration:STATe ON | OFF
This command determines whether (ON) or not (OFF) the current calibration data are taken into
consideration.
Example:
":CAL:STAT OFF"
Features:
*RST value:
-
SCPI:
conforming
Modes:
A, VA, BTS, MS
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......