
FSE SENSe Subsystem
1065.6016.12
6.165
E-16
SENSe:AVERage Subsystem
The SENSe:AVERage subsystem calculates the average of the data acquired. A new test result is
obtained from various successive measurements. The amount of test points and the axis reference of
the new result correspond to those of the original measurements.
COMMAND
PARAMETERS
UNIT
COMMENT
[SENSe<1|2>]
:AVERage
:COUNt
:AUTO
[:STATe]
:TYPE
<numeric_value>
<Boolean>
<Boolean>
MAXimum | SCALar
--
--
--
--
:[SENSe<1|2>:]AVERage:COUNt 0 to 32767
The command specifies the number of measurements which are combined.
Example:
":AVER:COUN 16"
Features:
*RST value:
0
SCPI:
conforming
Modes:
A. VA-D
:[SENSe<1|2>:]AVERage:COUNt:AUTO ON | OFF
AUTO ON selects a suitable number of :COUNt for the respective type of measurement.
Example:
":AVER:COUN:AUTO ON"
Features:
*RST value:
OFF
SCPI:
conforming
Modes:
A. VA-D
:[SENSe<1|2>:]AVERage[:STATe] ON | OFF
The command switches on or off the average function.
Example:
"
AVER OFF"
Features:
*RST value:
OFF
SCPI:
conforming
Modes:
A, VA, BTS, MS
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......