Status Reporting System
FSE
1065.6016.12
5.24
E-16
STATus:QUEStionable Register
This register comprises information about indefinite states which may occur if the unit is operated
without meeting the specifications. It can be queried by commands
STATus:QUEStionable:
CONDition?
and
STATus:QUEStionable[:EVENt]?
.
Table 5-5
Meaning of bits in STATus:QUEStionable register
Bit No.
Meaning
0
VOLTage
This bit is set if a questionable voltage occurs.
1
CURRent
This bit is set if a questionable current occurs.
2
TIME
This bit is set if a questionable time occurs.
3
POWer
This bit is set if a questionable power occurs (cf. also section "STATus:QUEStionable:POWerRegister")
4
TEMPerature
This bit is set if a questionable temperature occurs.
5
FREQuency
The bit is set if a frequency is questionable (cf. section "STATus:QUEStionable:FREQuency Register")
6
PHASe
The bit is set if a phase value is questionable.
7
MODulation
The bit is set if a modulation is performed questionably.
8
CALibration
The bit is set if a measurement is performed uncalibrated (=
^ label "UNCAL")
9
LIMit (unit-dependent)
This bit is set if a limit value is violated (see also section STATus:QUEStionable:LIMit Register)
10
LMARgin (unit-dependent)
This bit is set if a margin is violated (see also section STATus:QUEStionable:LMARgin Register)
11
SYNC (unit-dependent)
This bit is set if, during measurements with Option B7 (Signal Vector Analysis), the synchronization with
midamble or a successful search for bursts cannot be performed (see also STATus:QUEStionable:SYNC
Register)
12
ACPLimit (unit-dependent)
This bit is set if a limit for the adjacent channel power measurement is violated (see also section
STATus:QUEStionable:ACPLimit Register)
13
TRANsducer break
This bit is set when the limit of the transducer set subrange is attained.
14
COMMand Warning
This bit is set if the instrument ignores parameters when executing a command.
15
This bit is always 0.
The FSE supports bits 3, 5, 7, 8, 9, 10, 11, 12 and 13, bits 7 (MODulation) and 11 (SYNC) only with
option FSE-B7, Vector Signal Analysis’.
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......