DISPlay Subsystem FSE
1065.6016.12
6.94
E-16
:DISPlay[:WINDow<1|2>]:SELect
This command selects the active measurement window via the numeric suffix in WINDow . This way,
a switch is possible from FULL SCREEN A to FULL SCREEN B (see example).
Example:
":DISP:FORM SPLit"
":DISP:WIND2:SEL"
":DISP:FORM SINGle"
Features:
*RST value:
--
SCPI:
device-specific
Modes:
A, VA
This command is an event and has thus no query and no *RST value assigned.
:DISPlay[:WINDow<1|2>]:MINFo ON | OFF
This command switches the marker info list on the screen on or off.
Example:
":DISP:MINF ON"
Features:
*RST value:
OFF
SCPI:
device-specific
Modes:
A, VA, BTS, MS
:DISPlay[:WINDow<1|2>]:TEXT[:DATA] <string>
This command defines a comment (max. 50 characters) which can be displayed on the screen.
Example:
":DISP:TEXT "signal/noise power measurement"
Features:
*RST value:
"" (empty)
SCPI:
conforming
Modes:
A, VA, BTS, MS
The numeric suffix in WINDow<1|2> is not significant.
:DISPlay[:WINDow<1|2>]:TEXT:STATe ON | OFF
This command switches on or off the screen display of the comment.
Example:
":DISP:TEXT:STAT ON"
Features:
*RST value:
OFF
SCPI:
conforming
Modes:
A, VA, BTS, MS
The numeric suffix in WINDow<1|2> is not significant.
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......