FORMat Subsystem FSE
1065.6016.12
6.118
E-16
:FORMat:DEXPort:DSEParator POINt|COMMa
This command determines the decimal separator (decimal point or comma)for the output of the
measurement data in ASCII format. Thus, evaluation programms (i.e. MS-Exel) of different language
versions are supported.
Example:
":FORM:DEXP:DSEP POIN
Features:
*RST value:
POINt
SCPI:
device specific
Modes:
A, VA, BTS, MS
:FORMat:DEXPort:HEADer[:STATe] ON | OFF
This commad determines if the output file starts with a header (start frequency, sweep time, detector,
.. = ON) or not. OFF = only measurement data is output.
Example:
":FORM:DEXP:HEAD OFF
Features:
*RST value:
ON
SCPI:
device specific
Modes:
A, VA, BTS, MS
:FORMat:DEXPort:APPend[:STATe] ON | OFF
This commad determines if the output file is overwritten or the data is added to the end of the file.
Example:
":FORM:DEXP:APP OFF
Features:
*RSTvalue:
ON
SCPI:
device-specific
Modes:
A, VA, BTS, MS
:FORMat:DEXPort:COMMent <string>
This command defines a comment for the output file.
Example:
":FORM:DEXP:COMM ’ASCII EXPORT TRACE 1’"
Features:
*RST value:
„“
SCPI:
conforming
Modes:
A, VA, BTS, MS
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......