Brief Instructions
FSE
1065.6016.12
5.2
E-16
Brief Instructions
The short and simple operating sequence given below permits fast putting into operation of the
instrument and setting of its basic functions. As a prerequisite, the IEC-bus address, which is factory-set
to 20, must not have been changed.
1. Connect instrument and controller using IEC-bus cable.
2. Write and start the following program on the controller:
CALL IBFIND("DEV1", analyzer%)
’Open port to the instrument
CALL IBPAD(analyzer%, 20)
’Inform controller about instrument address
CALL IBWRT(analyzer%, "*RST;*CLS")
’Reset instrument
CALL IBWRT(analyzer%, ’FREQ:CENT 100MHz’)
’Set center frequency to 100 MHz
CALL IBWRT(analyzer%, ’FREQ:SPAN 10MHz’)
’Set span to 10 MHz
CALL IBWRT(analyzer%, ’DISP:TRAC:Y:RLEV -10dBm’)
’Set reference level to -10 dBm
The instrument now performs a sweep in the frequency range of 95 MHz to 105 MHz.
3. To return to manual control, press the LOCAL key at the front panel
Switchover to Remote Control
On power-on, the instrument is always in the manual operating state ("LOCAL" state) and can be
operated via the front panel.
It is switched to remote control ("REMOTE" state)
IEC-bus
as soon as it receives an addressed command from a controller.
RS-232
as soon as it receives the command ’@REM’ from a controller.
RSIB
as soon as it receives an addressed command from a controller.
During remote control, operation via the front panel is disabled. The instrument remains in the remote
state until it is reset to the manual state via the front panel or via remote control interfaces. Switching
from manual operation to remote control and vice versa does not affect the remaining instrument
settings.
Indications during Remote Control
Remote control mode is indicated by the LED "REMOTE" on the instrument’s front panel. In this mode
the softkeys, the function fields and the diagram labelling on the display are not shown.
Note:
Command
SYSTem:DISPlay:UPDate ON
activates all indications during remote control to
check the instrument settings.
Содержание 1065.6000.20
Страница 3: ......
Страница 5: ......
Страница 17: ......
Страница 57: ...Contents Description of Commands FSE 1065 6016 12 I 6 4 E 16...
Страница 379: ...Contents Programming Examples FSE 1065 6016 12 I 7 2 E 1...
Страница 391: ......
Страница 393: ...Contents Maintenance and Instrument Interfaces FSE 1065 6016 12 I 8 2 E 15...
Страница 419: ......
Страница 421: ...Contents Error Messages FSE 1065 6016 12 I 9 2 E 1...
Страница 429: ......