Command reference
R&S
®
ZNB/ZNBT
984
User Manual 1173.9163.02 ─ 62
Parameters:
<StimValue>
Stimulus values (unitless) confining the ripple limit range.
The unit is adjusted to the sweep type of the active channel
(
Range:
Virtually no restriction for ripple limit ranges.
*RST:
n/a (no ripple limit line defined after a reset)
Default unit: NN
Example:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:SEGM:STIM:STAR 1GHZ; STOP 2.5 GHZ; :
CALC:RIPP:SEGM:LIM 5
Change the range to a stimulus range between 1 GHz and 2.5
GHz and a limit of 5 dB.
CALC:RIPP:SEGM:STAT OFF
Disable the limit check in the modified stimulus range.
Manual operation:
See
CALCulate<Chn>:RIPPle:SOUNd[:STATe]
<Boolean>
Deprecated command
: Since version 3.45, generating sounds is no longer supported
by the analyzer firmware.
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded ripple limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - fail beep on or off.
*RST:
OFF
Example:
CALC:RIPP:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
CALCulate<Chn>:RIPPle:STATe
<Boolean>
Switches the ripple limit check for the active trace on or off.
Tip:
CALCulate<Chn>:RIPPle:SEGMent<Seg>[:STATe]
to switch the limit
check for a single ripple limit range on or off.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF
–
ripple limit check on or off.
*RST:
OFF
Example:
See
SCPI command reference