GUI reference
R&S
®
ZNB/ZNBT
462
User Manual 1173.9163.02 ─ 62
Transmission normalization and "One Path Two Ports" calibration types require two-
port (Through) characterization data for the cal unit. For current calibration units, how-
ever, these two-port characterizations are typically not part of the factory characteriza-
tion. If a tooltip indicates missing two-port characterization data, simply perform a new
characterization of your cal unit. In the first dialog of the "Characterization" wizard,
select "Take All OSM and Through" to make sure that the necessary two-port data is
acquired. See also
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Step 2: Connections
Defines the port assignments between the R&S
ZNB/ZNBT and the calibration unit.
Background information
Refer to
Chapter 4.5.5.6, "Multiple port assignments"
on page 191 for details on multi-
ple port assignments.
Figure 5-20: Multiple calibrations (and multiple port assignments)
Similar functionality as for the "SingleCal" version of the wizard (see
on page 453): select the adequate calibration in the "Cal Type"/"Ports" table to
display and edit the corresponding port assignments.
Back
Go back to
Start
.
If the configured port assignments are invalid, this action is disabled.
Cal softtool