Command reference
R&S
®
ZNB/ZNBT
1427
User Manual 1173.9163.02 ─ 62
SYSTem:TIME
<Hours>, <Minutes>, <Seconds>
The command queries or defines the instrument's current time setting.
The setting command requires administrator rights; refer to
Parameters:
<Hours>
Range:
0...23
<Minutes>
Range:
0...59
<Seconds>
Range:
0...59
Example:
SYST:TIME?
Response:
12,0,0
- it is precisely 12 pm.
SYSTem:TRESet[:STATe]
<Boolean>
Defines the behavior of the
INITiate<Ch>[:IMMediate][:DUMMy]
.
Parameters:
<Boolean>
OFF (0)
Previous trace data are preserved.
ON (1)
Previous trace data are deleted.
This is a global setting that is not affected by an instrument
reset. Factory default is OFF (0).
Manual operation:
See
"Restart: Set all Traces to 0"
SYSTem:TSLock
<Type>
Locks the touchscreen functionality of the R&S
ZNB/ZNBT, e.g. in order to prevent
inadvertent entries during remote control.
Parameters:
<Type>
OFF | DIAGrams | SCReen
OFF
–
touchscreen (e.g. the remote screen) remains active. You
can use the buttons in "Remote" softtool panels. If you switch
back to manual control, the drag and drop functions (e.g. for
markers) are still available.
DIAGrams
–
lock the drag and drop functions in diagrams .
SCReen
–
lock the entire screen including the remote screen
buttons.
*RST:
n/a (*RST does not affect the tochscreen lock set-
ting).
Example:
SYSTem:TSLock SCReen
Lock the entire screen.
Manual operation:
See
"Enabled / Lock Diagrams / Lock Screen"
SCPI command reference