GUI reference
R&S
®
ZNB/ZNBT
615
User Manual 1173.9163.02 ─ 62
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:
EMBedding<LogPt>[:STATe]
File Name 1
The ellipsis button in the "File Name 1" column is enabled if a 2-port data
selected.
When loading the touchstone file (
*.s2p
), the analyzer by default assumes odd ports
left (VNA side), even ports right (DUT side). However, it is also possible to
"Swap
Gates"
, instructing the analyzer to reinterpret the loaded S-parameters (e.g. S
12
-->
S
21
).
Note:
The loaded S-parameter traces are stored in the active recall set. Persisted
recall sets contain the full embedding and deembedding data so that they can be trans-
ferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:DIFFerential:EMBedding<LogPt>
5.13.2
Offset tab
Defines length offset parameters for each port.
Use the complementary dock widget to activate or deactivate length/loss compensation
for selected ports (see
Chapter 5.13.1.2, "Offset panel"
The marker function
function overwrites the offset parameters.
Background information
Refer to the following sections.
●
Chapter 4.6, "Offset parameters and de-/embedding"
●
Chapter 4.6.1.1, "Definition of offset parameters"
●
Chapter 4.6.1.3, "Auto Length"
●
Chapter 4.6.1.6, "Application and effect of offset parameters"
●
Chapter 4.6.1.7, "Offset parameters for balanced ports"
●
Chapter 4.6.1.5, "Fixture Compensation"
Offset Embed softtool