Command reference
R&S
®
ZNB/ZNBT
1201
User Manual 1173.9163.02 ─ 62
7.3.14.5
[SENSe:]CORRection:CKIT... with Labels
The following
[SENSe:]CORRection:CKIT...
commands identify the calibration kit
to be used by a combination of its <CalkitName> and <CalkitLabel>. Typically, the
serial number of the calibration kit serves as a calibration kit label. Due to their different
labels, the analyzer can handle several calibration kits with identical names.
[SENSe:]CORRection:CKIT:LCATalog?
.......................................................................... 1201
[SENSe:]CORRection:CKIT:LDELete
............................................................................. 1201
[SENSe:]CORRection:CKIT:LLABel
............................................................................... 1202
[SENSe:]CORRection:CKIT:LSELect
..............................................................................1202
[SENSe:]CORRection:CKIT:<ConnType>:LSELect
.......................................................... 1203
[SENSe:]CORRection:CKIT:STANdard:LCATalog?
...........................................................1204
[SENSe:]CORRection:CKIT:<StandardType>:WLABel
..................................................... 1204
[SENSe:]CORRection:CKIT:<OnePortStandardType>:WLABel:SDATa?
.............................1205
[SENSe:]CORRection:CKIT:<TwoPortStandard>:WLABel:SDATa?
.................................... 1205
[SENSe:]CORRection:CKIT:LCATalog?
[<ConnectorType>]
Returns a list of all cal kits and their labels for a given connector type or for all connec-
tor types.
Query parameters:
<ConnectorType>
Name of the connector type. Use
]CORRection:CONNection:CATalog?
names.
If omitted, the command returns the list of all cal kits andlabels.
Example:
See
[SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage:
Query only
[SENSe:]CORRection:CKIT:LDELete
<CalKitName>, <KitLabel>
Deletes an imported or user-defined cal kit which is identified by its cal kit name and
label.
Note:
It is not possible to modify or store predefined or ideal kits.
Setting parameters:
<CalKitName>
String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
<KitLabel>
String parameter containing the label of an imported or user-
defined calibration kit available on the analyzer.
Example:
See
[SENSe:]CORRection:CKIT:LLABel
Usage:
Setting only