Command reference
R&S
®
ZNB/ZNBT
1187
User Manual 1173.9163.02 ─ 62
Parameters:
<Boolean>
ON | OFF - enables or disables the automatic calculation of the
sweep average over the specified number of sweeps
(
*RST:
ON
Example:
See
Manual operation:
See
[SENSe<Ch>:]AVERage:CLEar
Starts a new average cycle, clearing all previous results and thus eliminating their
effect on the new cycle.
Suffix:
<Ch>
.
Channel number
Example:
SENS1:AVER:COUN 15; :AVER ON
Set the average factor for channel 1 to 15 (the mnemonic
SENS1
can be omitted) and enable the sweep average.
AVER:COUN 5; CLE
Reduce the average factor and restart the average.
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]AVERage:COUNt
<AverageFactor>
Defines the number of consecutive sweeps to be combined for the sweep average
("Factor").
Suffix:
<Ch>
.
Channel number
Parameters:
<AverageFactor>
Sweep average factor
Range:
1 to 1000
*RST:
10
Example:
See
Manual operation:
See
[SENSe<Ch>:]AVERage:MODE
<Mode>
Suffix:
<Ch>
.
Parameters:
<Mode>
AUTO | FLATten | REDuce | MOVing
SCPI command reference