Command reference
R&S
®
ZNB/ZNBT
1189
User Manual 1173.9163.02 ─ 62
Manual operation:
See
[SENSe<Ch>:]BANDwidth[:RESolution]:SELect
<Selectivity>
[SENSe<Ch>:]BWIDth[:RESolution]:SELect
<Selectivity>
Defines the selectivity of the IF filter for an unsegmented sweep. The value is also
used for all segments of a segmented sweep, provided that separate selectivity setting
is disabled
(
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol
OFF
).
Suffix:
<Ch>
.
Channel number
Parameters:
<Selectivity>
NORMal | MEDium | HIGH
NORMal – IF filter with normal selectivity and short settling time.
MEDium – IF filter with steeper edges and longer settling time.
HIGH – IF filter with highest selectivity but longest settling time.
*RST:
NORMal
Example:
See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:
Manual operation:
See
7.3.14.3
[SENSe:]CDLL...
Adds, removes and configures
These commands allow you to add/remove external DLLs to/from the firmware. It gives
access to their configuration, tasks, and traces.
...........................................................................................1190
[SENSe<Ch>:]CDLL:PERManent:ADDitional
.................................................................. 1190
[SENSe<Ch>:]CDLL:PERManent[:STATe]
.......................................................................1191
[SENSe<Ch>:]CDLL:PERManent:TASK
..........................................................................1191
................................................................................................1191
......................................................................................... 1192
.............................................................................1192
[SENSe:]CDLL:ADD
<DllPath>
Tells the analyzer firmware to load a custom (external) DLL.
The DLL and its accompanying files must be packed in a zip archive. It will be loaded
on every subsequent firmware start, until it is removed using
.
SCPI command reference