GUI reference
R&S
®
ZNB/ZNBT
497
User Manual 1173.9163.02 ─ 62
a) Acquire error correction data for the required ports and port pairs, and the
required standards.
b) Acquire source power calibration data at the "Power Port".
c) Finally, decide whether to apply the resulting calibration.
Background and related information
●
on page 199 for background information.
●
Chapter 5.11.1.11, "SMARTerCal wizard (manual, MultiCal)"
the "MultiCal" variant of the wizard.
●
to perform an additional source flatness calibration (see
Chapter 4.5.7.3, "Combining SMARTerCal with scalar power calibration"
●
If the active channel is already calibrated, the wizard attempts to load the calibra-
tion. If the calibration setup is unchanged and sweep data are available, the exist-
ing system error correction can be optimized without repeating the measurement of
all standards.
●
When you apply the acquired SMARTerCal, the active calibration is replaced and
discarded.
To persist any kind of calibration, you can transfer it to the cal "Pool" using the
.
●
The active system error correction data can be read (and modified) using the
remote control command
If external switch matrices are involved, use
Step 1: Ports and Type
Selects the calibrated analyzer ports, the calibration type and specifies the power
meter connection.
Background information
Refer to
Chapter 4.5.7.2, "Calibration types"
Cal softtool