Command reference
R&S
®
ZNB/ZNBT
1264
User Manual 1173.9163.02 ─ 62
<PhyPt>
Port number. This suffix is ignored; the analyzer calibrates the
port that is selected as a receive port for the intermodulation
measurement
(
[SENSe<Ch>:]FREQuency:IMODulation:RECeiver
).
Setting parameters:
<CalOnlyPortFreq>
OFF | 0
The reference receiver at port <CalPort> is calibrated for all fre-
quencies that are relevant for any of the ports.
ON | 1
Only those frequencies are calibrated that are relevant for this
port. This results in shorter calibration times.
*RST:
0
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:POWer<PhyPt>[:STATe]
<Boolean>
Enables or disables the receiver power calibration for channel
<Ch>
and for the
received waves b
<PhyPt>
. The setting command is disabled unless the received waves
have been power calibrated (
[SENSe<Ch>:]CORRection:POWer<PhyPt>:
BWAVe, ...
). The query always returns a result.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the receiver power calibration
for the received waves b
<PhyPt>
.
*RST:
OFF
SCPI command reference