GUI reference
R&S
®
ZNB/ZNBT
447
User Manual 1173.9163.02 ─ 62
Start Auto Cal
Performs a fully automatic system error calibration for the active channel, using the
active calibration unit with its factory calibration and auto-detection of ports.
If one port is active, an OSM calibration is performed. If two ports are active, an UOSM
calibration is performed.
Calibration – Start... (Cal Unit)
Depending on the
Multiple Cal in Calibration Wizard
setting, the "Start... (Cal Unit)" but-
ton either opens the
Calibration Unit wizard (MultiCal)
.
Calibration – Start... (Manual)
Depending on the
Multiple Cal in Calibration Wizard
setting, the "Start... (Cal Unit)" but-
ton either opens the
Calibration Setting wizard (Multi-
Calibration – Repeat...
Re-enters the data acquisition step of the current calibration, restoring the related
channel settings. This provides a convenient way to repeat or correct an existing cali-
bration without necessarily repeating all measurements.
Scalar Power Cal – Power Cal...
.
If the active recall set contains several channels, an "Info" box lets you decide whether
the R&S
ZNB/ZNBT acquires calibration data for the active channel or for all channels.
SMARTerCal – Start... (Cal Unit)
Depending on the
Multiple Cal in Calibration Wizard
setting, the "Start... (Cal Unit)" but-
ton either opens the
or the
.
SMARTerCal – Start... (Manual)
Depending on the
Multiple Cal in Calibration Wizard
setting, the "Start... (Cal Unit)" but-
ton either opens the
or the
.
SMARTerCal – Repeat...
Reenters the data acquisition step of the current SMARTerCal, restoring the related
channel settings. This function provides a convenient way to repeat or correct an exist-
ing SMARTerCal without necessarily repeating all measurements.
Cal Validation ...
that allows you to set up and run a
.
5.11.1.2
Channel Selection for Calibration dialog
If you initiate a calibration and there is more than one channel in the current recall set,
you are prompted to select the channels to be calibrated:
Cal softtool