Command reference
R&S
®
ZNB/ZNBT
904
User Manual 1173.9163.02 ─ 62
CALCulate:FMODel:EZD<Ph_pt>:COUPon[:STATe]
<Boolean>
For a fixture modeling with the EZD tool, this command allows to specify the ports to
which the test coupon is connected.
This has to be defined before measuring the test coupon (using
).
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean>
ON (1)
if the test coupon is connected to port <Ph_pt>
OFF (0)
otherwise
Options:
R&S
ZNB/ZNBT-K210
Manual operation:
See
CALCulate:FMODel:EZD:DCEXtrapolat
<Boolean>
Defines whether the Touchstone de-embedding file generated by the EZD tool contains
data for DC.
Parameters:
<Boolean>
ON (1)
The EZD tool adds (extrapolated) S-parameter data for f = 0.
OFF (0)
The frequencies only reach down to the start frequency of the
instrument sweep.
Options:
R&S
ZNB/ZNBT-K210
Manual operation:
See
CALCulate:FMODel:EZD:DUT:MEASure
For a fixture modeling with the EZD tool, this command allows to measure "DUT + Test
Fixture" at the active ports (see
CALCulate:FMODel:EZD<Ph_pt>:DUT[:STATe]
)
Usage:
Event
Options:
R&S
ZNB/ZNBT-K210
Manual operation:
See
CALCulate:FMODel:EZD:DUT:MEASure:FILename
<String>
For a fixture modeling with the EZD tool, this command loads the "DUT + Test Fixture"
properties from a Touchstone file.
SCPI command reference