Command reference
R&S
®
ZNB/ZNBT
1168
User Manual 1173.9163.02 ─ 62
<CalGroupFile>
String parameter to specify the name of the created cal group
file. There is no need to specify the directory path and file exten-
sion; the analyzer uses the default cal pool directory
C:\Users\Public\Documents\Rohde-Schwarz\Vna
\Calibration\Data
and a
*.cal
extension.
Example:
See
Usage:
Setting only
Manual operation:
See
"Add / Add All... / Replace / Apply / Apply to All"
MMEMory:STORe:CORRection:TCOefficient<Ch>
<PmclFile>
Saves two-port transmission coefficients of channel <Ch> to a power meter correction
list file.
Suffix:
<Ch>
.
Channel number
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using
SOURce<Ch>:POWer<PhyPt>:CORRection:
Setting parameters:
<PmclFile>
String parameter specifying the name and directory of the cre-
ated power meter correction list file. The file extension
*.pmcl
is mandatory.
If no path is specified, the analyzer searches the current direc-
tory, to be queried with
*RST:
n/a
Example:
See
SOURce:POWer:CORRection:TCOefficient[:STATe]
Usage:
Setting only
Manual operation:
See
MMEMory:STORe:EYE:MASK
<TraceName>, <TraceFile>
Stores a user-defined eye mask to a 7bit ASCII file.