Command reference
R&S
®
ZNB/ZNBT
1316
User Manual 1173.9163.02 ─ 62
Parameters:
<Boolean>
ON
–
use independent delay settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl
OFF
–
reset the delay in all sweep segments to the delay for
unsegmented sweeps, defined via
The parameter is automatically switched to ON when a meas.
delay time is entered using
*RST:
OFF
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus 0 s meas. delay.
SEGM:SWE:DWELl 0.1
Increase the meas. delay to 0.1 s.
SEGM:SWE:DWELl:CONT OFF
Couple the meas. delay in all segments and reset the delay in
segment no. 1 to the initial value of 0 s.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:GENeration
<arg0>
Sets/gets the segment-specific sweep modes for segmented linear frequency sweeps.
Suffix:
<Ch>
.
Channel number
<Seg>
Number of an existing sweep segment.
Parameters:
<arg0>
STEPped | ANALog
Stepped mode or swept mode (=ANALog).
Example:
*RST
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Creates segment 1 with the given parameters and the default
frequency sweep mode
STEP
.
SEGM:INS 3MHZ, 4.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ, POS, NORM, AUTO, ANAL
Inserts a new segment 1 with the given parameters, moving the
previous segment 1 to segment 2. Hence
SENSe:SEGMent1:SWEep:GENeration?
returns
ANAL
and
SENSe:SEGMent2:SWEep:GENeration?
returns
STEP
.
Manual operation:
See
SCPI command reference