- 50 -
Caution: For information on how to design printouts go to section 14.5.
Air buoyancy compensation
Parameters allowing you to switch on the compensation and to enter weighed sample
density data and air density data.
CAUTION! The function is valid for weighing mode exclusively.
For information on how to set air buoyancy compensation go to the previous section.
15.8. MINIMUM SAMPLE WEIGHT
The weighing mode settings comprise <Minimum sample weight> function. In order to use the
function it is necessary to enter data related to minimum sample weight determination methods and
minimum sample weight values for a given method. The data must be entered to < Minimum sample
weight> database. The database is empty in standard balance version.
Determining minimum sample weight and adding data to the database: </Minimum sample
weights> database can be carried out by an authorized “RADWAG” personnel exclusively.
Should you use the minimum sample weight function, and need data on minimal sample weight to be
entered to balance menu, please contact the local representative of “RADWAG” company.
An authorized RADWAG representative determines the minimum sample weight depending on
packaging mass. The process is carried out with mass standards, in balance place of use and in
accordance with the requirements of the accepted quality system. The obtained values are entered to
balance software: <Minimum sample weights> database.
A minimum sample weight determining method may have a few tare values defined, along with
assigned values of minimum sample weights, expiry dates of the carried out measurements and
entered data. The settings are not to be edited by a balance user.
Using <Minimum sample weight> function guarantees that the results of completed weighing
processes are within set tolerance limits, and they comply with the requirements of the quality system
accepted by an organization.
CAUTION! The function is valid for weighing mode exclusively.
PARAMETER VALUES:
•
METHOD
Option allowing you to designate used quality standard. Press METHOD field to open a window
with a list of database-stored methods by means of which the minimum sample weights were
determined. In order to add a new method you have to go to <Databases /Minimum sample
weights>
•
MODE
Lock
– select in order to make the balance display pictograms informing you on status of
weighed mass (weighed mass below/above minimum sample weight) and to disable accepting
the measurement result if it is below the minimum sample weight limit
Warn –
select in order to make the balance display pictograms informing you on status of
weighed mass (weighed mass below/above minimum sample weight). You can accept the
measurement result if it is below the minimum sample weight limit, mind that on a printout it will
be preceded by an asterisk (*).
CAUTION!
For information on how to enter a method for minimum sample weight go to section 30.14 <Databases
/Minimum sample weights>. New minimum sample weight values can be entered by an authorized “RADWAG”
personnel exclusively.
Weighing with use of <MINIMUM SAMPLE WEIGHT> function
With <Minimum sample weight> function on you are informed whether the measurement is out of
minimum sample weight limit determined for a specific weighing range. To turn the function on go to
the weighing mode settings.
Procedure:
1. Press grey workspace.
2. Settings, Keys, Information, Printouts and Profiles submenus are displayed.
Содержание UYA 4Y