- 96 -
•
weighing T procedure only – weighing A value is <0>,
•
weighing T+A procedure – tare and weighing A values is mass.
You can print weighed sample data, to do it press a printer pictogram located in the top bar. In order
to preview data on weighing A, press <Weighing A> field:
If a given series already contains samples for which weighing A (initial mass) has been carried out
you can proceed to carrying out weighing B (final mass).
Weighing B
•
On pressing Weighing B button, the balance searches samples list in order to find the first
sample for which weighing <B> is still to be carried out. If there are no such samples the
balance displays a message informing you that the operation is disabled.
•
If the operation is enabled then new data on initiated process is displayed in the grey
workspace.
Caution:
Next to <Weighing B> caption you can see digits 1/3 where <1> means the first measuring cycle of
type B, and <3> means that <REPETITIONS QUANTITY> parameter is set to <3> parameter (3
cycles).
The software prompts you to carry out the <FIRST> cycle, out of <3> cycles, for all samples in the
series for which the measurement is enabled (it is required that weighing A is already completed for
the sample).
•
You are guided through the process by means of prompts visible either in the workspace or
bottom bar.
•
If tare is assigned to a sample, then its value is displayed with ‘-‘ sign. Load the weighing pan
with a sample (if tare is assigned load the sample in a container) and press <
> button. For
<Ambient condition> parameter set to <VALUE> parameter a window for entering temperature
and humidity values is displayed. For <Ambient condition> parameter set to <ONLINE>
parameter a window for confirmation of values taken from ambient conditions module is
displayed.
Caution:
If <Ambient conditions> parameter for the differential weighing mode is set to <No> parameter the
software skips this part of the process.
•
Press <
> button to confirm ambient conditions values, differential weighing home screen is
displayed, you are prompted to unload the weighing pan.
•
Unload the weighing pan and press <
> button. The balance searches samples list again in
order to find the next sample in the series. Follow the above procedure and weigh the next
sample.
•
In order to abort the process press <
> button.
On completing measurements of final mass for the samples in a series, you can check the results in
the series database.
To check the results, enter the database, select a series, then select samples and a specific sample
for which <Weighing B> has been carried out.