- 133 -
sample weight settings.
Thresholds
– option that enables entering data on minimum sample
weight mass and range of packaging mass (tare) with
applicable criteria:
Tare
– maximum value of tare for which the minimum sample weight is valid.
The software enables entering three characteristic quantities: 0.000g, optional mass from balance
measuring range and Max balance measuring range (see example below).
Minimum mass –
the value of minimum mass that is determined for a balance in its place of use in
accordance with appropriate methods.
Example no.1 XA 220.4Y balance, d=0.0001 g:
The following values were determined for this balance:
No.
Tare value
Minimum
sample
weight
Means of operation
1
0.0000 g
0.1000 g
The minimum sample weight refers to all net
weighments that are weighed without a packaging
(the <TARE> button is not applied)
2
10.0000 g
1.0000 g
The minimum sample weight refers to all net
weighments that are weighed in a packaging that
mass ranges from 0.0001 g to 10.0000 g inclusive
(the <TARE> button applied)
3
50.0000 g
2.5000 g
The minimum sample weight refers to all net
weighments that are weighed in a packaging that
mass ranges from 10.0001 g to 50.0000 g inclusive
(the <TARE> button applied)
4
200.0000 g
4.0000 g
The minimum sample weight refers to all net
weighments that are weighed in a packaging that
mass ranges from 50.0001 g to 200.0000 g inclusive
(the <TARE> button applied)
Example no. 2 XA 220.4Y balance:
No.
Tare value
Minimum
sample
weight
Means of operation
1
220.0000 g
0.5000 g
The minimum sample weight refers to all net
weighments that are weighed in a packaging with
optional mass from full measuring range of a balance
(the <TARE> button applied), as well as on weighing
samples without the packaging (the <TARE> button is
not applied).
Example no. 3 XA 220.4Y balance:
No.
Tare value
Minimum
sample
weight
Means of operation
1
0.0000 g
0.2500 g
The minimum sample weight refers to all net
weighments that are weighed without a packaging
(the <TARE> button is not applied).
Minimum sample weight is valid only for weighing
samples without packaging. If taring option is applied,
then the software disables the pictogram informing on
minimum sample weight in use, and renders the
weighment as carried out with tare applied (minimum
sample weight is not determined).
You can preview entered data, but data editing is disabled.