- 30 -
In order to add a new position, click
button, wait for
a window with data on planned balance adjustment to
open.
Select task: automatic adjustment (internal) or external
calibration.
For automatic adjustment option, enter data on the
adjustment and its schedule.
For external calibration option, enter data on the
calibration, mass standard used for the calibration
performance and its schedule.
When all necessary data has been introduced, go back
to the previous window. The window features newly
added position relating to planned balance adjustment
procedure.
Upon entering all data exit the menu.
Now all the adjustments are carried out automatically as planned and with specified time intervals.