- 37 -
Value release
The parameter has been designed to determine when the stability marker is to
be displayed.
There are three different settings for value release parameter:
•
fast
•
fast + reliable /recommended/
•
reliable
CAUTION!
Both parameters, filter and value release, determine how long it takes to obtain stable
result.
Autozero function
The function has been designed to enable automatically carried out monitoring
and correction of zero indication.
If the function is enabled, the following measurement results are compared to
each other in constant time intervals. If the results differ less than declared
AUTOZERO range, e.g. 1 division, the balance is automatically zeroed, and the
markers of stable indication
and precise zero
are displayed.
Active AUTOZERO means, that each measurement starts from the precise zero
point. There are, however, cases where the function may disturb the weighing
process. For instance during very slow load placing or pouring onto the
balance’s weighing pan. In such case, the correcting system of zero indication
may also correct the actual indication of a load placed on the weighing pan.
Accessible settings:
NO
- autozero function disabled
YES
- autozero function enabled
Autozero function: Dosing
The function has been designed to set autozero to default value selected for
dosing.
Accessible settings:
NO
- autozero operation is automatically disabled on entering dosing mode
YES
- autozero operation is automatically enabled on entering dosing mode
Last digit
The function has been designed to enable/disable visibility of the last decimal
place of measurement result.
Accessible settings
:
ALWAYS
: al the digits visible
NEVER
: last digit invisible
WHEN STABLE
: last digit visible only for a stable measurement
Ambient conditions
The parameter refers to ambient conditions. By default it is set to value
Stable.
For unstable ambient conditions it is advisable to set the parameter to value
Unstable
.
Stable
setting results with faster operation, i.e. weighing takes much less time
than in case of setting to value
Unstable
.
Accessible settings
:
Unstable
Stable
13.2.4. Measuring Units
For a selected profile you can declare start unit, supplementary unit and two custom measuring units.