- 169 -
Format:
LOGOUT CR LF
Response options:
LOGOUT OK CR LF
- command understood, user is logged out
ES CR LF
- command not recognized (error in format)
Value release
Format:
SS CR LF
Response options:
SS_OK CR LF
- command understood and in progress
Command’s function is similar to function of PRINT button located on an overlay,
Internal adjustment
Format:
IC CR LF
Response options:
IC_A CR LF
- command understood and in progress
IC_D CR LF
- calibration completed
IC_A CR LF
- command understood and in progress
IC_E CR LF
- time limit exceeded while waiting for a stable measurement result
IC_I CR LF
- command understood but not accessible at this moment
Disable automatic internal adjustment of the balance
Format:
IC1 CR LF
Response options:
IC1_I CR LF
- command understood but not accessible at this moment
IC1_E CR LF
- operation disabled, e.g. for verified balance
IC1_OK CR LF
- command carried out
Operation is disabled for verified balances.
For non-verified balances the command inhibits internal calibration until it is enabled via IC0
command or until the balance is turned off. The command does not modify settings specifying
calibration start.
Enable automatic internal adjustment of the balance
Format:
IC0 CR LF
Response options:
IC0_I CR LF
- command understood but not accessible at this moment
IC0_OK CR LF
- command carried out
Operation is disabled for verified balances.
Lock balance keypad
Format:
K1 CR LF
Response options:
K1_I CR LF
- command understood but not accessible at this moment
K1_OK CR LF
- command carried out
Command locks the balance keypad (proximity sensors, touch panel) until the moment of
turning the balance off or until sending K0 command
Unlock balance keypad
Format:
K0 CR LF
Response options:
K0_I CR LF
- command understood but not accessible at this moment
K0_OK CR LF
- command carried out
Activate sound signal
Command overview: command activates BEEP sound signal for a specified amount of time
Format:
BP_CZAS CR LF
Response options:
BP_OK CR LF
– command carried out, BEEP sound signal activated
BP_E” CR LF
– no parameter or incorrect format
BP_I CR LF
– command understood but not accessible at this moment
TIME
-parameter, decimal value specifying how long the sound lasts,
Содержание UYA 4Y
Страница 8: ...8...
Страница 9: ...9 1 GENERAL INFORMATION 1 1 DIMENSIONS Ultra microbalances and microbalances...
Страница 10: ...10 MYB 4Y Microbalances Indicator XA 4Y series balances...
Страница 11: ...11 XA 4Y F series balances...
Страница 195: ...195...
Страница 196: ...196...