- 175 -
Format of mass printout:
1
2
3
4 -12
13
14
15
16
17
18
stability
marker
space
character
mass
space
unit
CR
LF
Stability marker
[space] if stable
[?] if unstable
[!] if air buoyancy compensation mode is enabled
[^] if error of exceeding range to + occurs
[v] if error of exceeding range to – occurs
Character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point and right justification
Unit
3 characters with left justification
Example:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF
– a printout generated from a balance on pressing
key, according to settings of <Weighing printout template>:
N (measurements
quantity)
NO
Universal variable 1 … 5
NO
Date
NO
Net
NO
Time
NO
Tare
NO
Level status
NO
Gross
NO
Customer
NO
Current result
NO
Warehouse
NO
Supplementary unit
NO
Product
NO
Mass
YES
Packaging
NO
Non-standard printout
NO
39. CONNECTING PERIPHERALS
Y series balances can cooperate with the following peripheral devices:
•
Computer,
•
Receipt printer: KAFKA, EPSON,
•
PCL printer,
•
Additional display,
•
Barcode scanner,
•
An optional peripheral device operating in communication protocol ASCII.
40. ERROR MESSAGES
Err2
Value beyond zero range
Err3
Value beyond tare range
Err8
Taring / Zeroing operation time exceeded
-NULL-
Zero value from converter
-FULL-
Measurement range (Max. capacity) exceeded
-LH-
Start mass error