- 29 -
•
using displayed menu, select appropriate time interval (given in hours) which is a time gap
elapsing from the last carried out internal automatic adjustment until activation of the following
automatic internal adjustment.
CAUTION!
Parameter settings can be modified only for balances that are not a subject to conformity assessment
(verification).
10.7. SCHEDULED BALANCE ADJUSTMENT
<
Scheduled adjustments> parameter allows you to specify precisely time for adjustment
performance and interval between successive adjustments. The parameter is independent from
automatic adjustment and criteria that trigger it (time, temperature). You can plan when internal and
external adjustment shall be performed. In order to design schedule for external adjustment, it is
necessary to record into balance memory mass standards for adjustment performance.
Parameter values:
1. Enter data of mass standards intended for external adjustment performance.
2. Enter user menu, select <Adjustment> parameter group and find <
Mass standards>
parameter, enter necessary data:
Enter user menu, select <Adjustment> parameter
group and find <Scheduled adjustments> parameter.
Enter <Scheduled adjustments> parameter, wait for
‘Scheduled adjustments’ window to open, add
adjustment procedures. Remember, only Administrator
can add new positions.