- 168 -
U
H
space
mass
space
unit
space
CR
LF
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
Set mass value of a single item (only for PARTS COUNTING)
Format:
SM_XXXXX CR LF
, where:
_ -
space
,
XXXXX
– mass format
Response options:
SM_OK CR LF
- command carried out
SM_I CR LF
- command understood but not accessible at this moment (e.g. mode
other than PARTS COUNTING)
ES CR LF
- command not recognised (mass format incorrect)
Set target mass value (e.g. for DOSING)
Format:
TV_XXXXX CR LF
, where:
_ -
space
,
XXXXX
- mass format
Response options:
TV_OK CR LF
- command carried out
TV_I CR LF
- command understood but not accessible at this moment (e.g. mode
other than DOSING)
ES CR LF
- command not recognised (mass format incorrect)
Set reference mass value (e.g. for DEVIATIONS)
Format:
RM_XXXXX CR LF
, where:
_ -
space,
XXXXX
– mass format
Accessible responses:
RM_OK CR LF
RM_I CR LF
- command carried out
- command understood but not accessible at this moment (e.g.: oth
working mode, not PERCENT WEIGHING)
ES CR LF
- command not recognized (incorrect mass format)
Give balance serial number
Format:
NB CR LF
Response options:
NB_A_”x” CR LF
- command understood, response: serial number
NB_I CR LF
- command understood but not accessible at this moment
x – serial number of the device (inserted in between inverted commas)
An example:
Command:
NB CR LF – return serial number
Response:
NB_A_”1234567” – serial number of the device – ”1234567”
Set balance mode
Format:
PROFILE_Name CR LF
where:
_ -
space (Name must be accordant with the name form entered in a balance –
lowercase and uppercase letters
)
Response options:
PROFILE OK CR LF
- command understood, new PROFILE activated
LOGIN ERRROR CR LF
- command understood, an error in PROFILE name, log in failed
ES CR LF
- command not recognized (error in format)
User log in
Format:
LOGIN Name, Password CR LF
(Name and password have to be inserted as they a
set in the balance – small and capital letters)
Response options:
LOGIN OK CR LF
- command understood, new user is logged in
LOGIN ERRROR CR LF
- command understood, an error in name or password
occurred, log in failed
ES CR LF
- command not recognized (error in format)
User log out