- 157 -
settings providing optimal measurement (the lowest value of
time and deviation product)
current settings for filters
Measurement results:
•
Filter type
•
Setting of
<Value release>
parameter
•
Repeatability of indications expressed as standard deviation
•
Average time of measurement stabilization
Example:
You can quickly modify filters settings and settings for results approval.
Procedure:
1. Select a group of results for particular settings that are to be saved.
2. Press
<Activate>.
Report template:
-------- Autotest Filter: Report --------
Balance type
XA 4Y
Balance ID
442566
User
Hubert
Software rev.
NL1.6.5 S
Date
2015.05.07
Time
09:34:48
----------------------------------------
Reading unit
0.0001/0.0001 g
Internal weight mass
148.9390 g