- 15 -
CAUTION: Make sure there is no protective sticker 1 while carrying measurement (see the above
diagram).
Install components following the above diagram depending on balance’s type.
A – standard (UYA and MYA)
2. Anti-draft chamber
3. Weighing pan
4. Glass lid for the weighing pan
B – MYA 0.8/3.4Y microbalances
Balances are equipped with standard weighing
pan (A) and weighing pan for filters.
12. Bottom ring
9. Weighing pan for filters
4. Glass lid for the weighing pan
C – balance for filters
12. Bottom ring
10. Weighing pan for filters
11. Standard weighing pan
D – pipettes calibration balance
2. Anti-draft chamber
3. Weighing pan
5. Glass vessel
6. Evaporation ring
7. Glass lid for the weighing pan, with an
opening
8. Additional glass lid
4. Glass lid for the weighing pan
Pipette calibration:
On balance activation replace the weighing pan with weighing pan with glass vessel (5). When the
indication is stable press TARE button.