- 155 -
3. Use finger or a thin (but not sharp edged) and soft stylus to press and hold a display area
occupied by a cross. After pressing the fifth cross in order, accept the changes by pressing
key.
Caution:
The process can be aborted by pressing Esc key on an external keyboard connected to the balance’s terminal.
34.8. LEVEL CONTROL
4Y series balance features an Automatic Level Control system. In case of non-verified instruments,
you can define the operation of Automatic Level Control system. In case of verified instruments, the
settings are hidden and the balance operates in accordance with factory settings, i.e.
<Active with
lock>
. Weighing process is possible only if balance is properly leveled.
Procedure:
1. Enter
<
Miscellaneous>
submenu.
2. Select
<
Level control>
parameter, an editing window is displayed.
3. Select one of available options:
•
None – level indicator is not displayed, the balance does not control level status,
•
Active – level indicator is displayed, the balance alerts level change by changing colour of
level indicator (green
→
level status OK., red
→
level status lost)
•
Active with lock – level indicator is displayed, the balance alerts level change by changing
colour of level indicator (green
→
level status OK., red
→
level status lost). If level
indicator is red, the display also shows an alert – no Level - , and blocks weighing
function until level status is restored.
Caution:
Balance levelling is described in section 15.3 of this user manual.
34.9. SENSORS SENSITIVITY
The parameter is adjustable between 0 – 9, and it determines the distance from
which the IR proximity sensors react.
As standard, the sensitivity is set within the range of 5 to 7.
Procedure:
1. Enter
<
Miscellaneous>
submenu.
2. Select
<Sensors sensitivity>
parameter, an editing window is displayed.
3. Select one of available options. The selection causes immediate returning to the menu
window.
34.10. IR SENSORS TIMEOUT
Parameter for setting timeout of proximity sensors expressed in [ms]. The default value is <0>.
When using option of motion recognition you must remember about setting
<IR sensors
timeout>
parameter to value
<500ms>.
Procedure:
1. Go to
<
Miscellaneous>
submenu.
2. Select
<Proximity sensors delay>
parameter, an on-screen keyboard is displayed.
3. Enter value
<500>
.
4. Confirm. The software returns to menu window.