To configure barcode scanner follow instruction below:
„
/
Peripherals /
Barcode scanner”
.
Caution:
<
Communication> submenu requires setting baud rate to compatible with the one used by barcode
scanner (default 9600b/s). For detailed description of communicating the balance with a barcode scanner go to
APPENDIX E of this user manual.
32.3.1. Barcode Scanner’s Port
Procedure:
1. Open <
Peripherals> group of parameters.
2. Press
<
Barcode scanner>.
3. Select suitable option of those presented below.
Ports:
•
RS 232 (COM1),
•
RS 232 (COM2),
•
USB
32.3.2. Prefix/Suffix
You can edit
<
Prefix>
and
<
Suffix>
to synchronize balance’s program with barcode
scanner.
Caution
In RADWAG balances prefix contains character (byte) presented as 01 hexadecimal, whereas suffix contains
character (byte) 0D hexadecimal. For detailed description of communicating the balance with a barcode
scanner go to APPENDIX E of this user manual.
Procedure:
1. Open
<
Barcode scanner>
submenu.
2. Select
<
Prefix>
parameter and enter hexadecimal value using on-screen keyboard.
3. Press
to confirm changes.
4. Select
<
Suffix>
parameter and enter hexadecimal value using on-screen keyboard.
5. Press
to confirm changes.
32.3.3. Field Selection
You can customize field selection in databases after which searching is carried out.
Procedure:
1. Open <
Peripherals> group of parameters.
2. Select
<
Barcode scanner/
Fields selection>
.
3. Following list is displayed:
Product
Series
User
Universal variable 1