i-D33P75SD Bench Scale 

Quick Start Guide 

 

 

 

 
 
 
 

Содержание i-D33P75SD

Страница 1: ...i D33P75SD Bench Scale Quick Start Guide ...

Страница 2: ......

Страница 3: ...ad to equipment damage if not avoided NOTE For useful information about the product Warning Symbols General hazard Explosion hazard Electrical shock hazard Safety Precautions CAUTION Read all safety warnings before installing making connections or servicing this equipment Failure to comply with these warnings could result in personal injury and or property damage Retain all instructions for future...

Страница 4: ...and qualified personnel Remove all power connections to the unit before opening 2 INSTALLATION 2 1 External Connections RS232 Connection Connect the optional RS232 cable to the RS232 connector on the rear housing of the i DT33P indicator Please refer to table 2 1 for the definition of each pin if needed Rechargeable Battery Power Option Allow the battery to charge for 12 hours before using the sca...

Страница 5: ... power off Units Changes the weighing unit Mode Changes the application mode The i D33P75SD scale should be used in Weigh mode for materials testing Menu Enters the user menu Shows a tare value in application modes Menu Function Short Press Yes Accepts the current setting on the display or selects a sub menu or menu item No Advances to the next menu or menu item Rejects the current setting on the ...

Страница 6: ...ibration point is established with no weight on the scale Use this calibration method to adjust for a different static load without affecting the span or linearity calibration Calibration procedure when the calibration unit is kg 1 Long press the Menu button until you see mM E N U 2 Release the button and wait for the display to show C A L Press the Yes button 3 The display shows ZErO Press the Ye...

Страница 7: ...on the display 6 Place a calibration mass of the specified weight on the pan and press the Yes button 7 The display shows C and then DONE when the calibration is finished 8 Then the display shows Lin Press the Exit button to exit Note If calibration has failed CAL E is displayed for 3 seconds and the previous calibration data is restored If after waiting for 40 seconds the calibration is still not...

Страница 8: ...ecified weight on the pan and press the Yes button The display shows C 8 The second calibration point and the unit are flashing on the display based on the capacity and unit you set in the setup menu For example 030 000 kg Note If calibration has failed CAL E is displayed for 3 seconds and the previous calibration data is restored If after waiting for 40 seconds the calibration is still not succes...

Страница 9: ...mmonia or abrasive cleaning agents 5 2 Troubleshooting For technical issues contact an Authorized Ohaus Service Agent Please visit our website www ohaus com to locate the OHAUS office nearest you 6 TECHNICAL DATA 6 1 Specifications Equipment Ratings Altitude 2 000m Operating temperature 10 C to 40 C Humidity Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relativ...

Страница 10: ...ith backlight off Rechargeable battery optional up to 75 hours continuous use with backlight off Operating Temperature Range 10 C to 40 C 14 F to 104 F Indicator to Base Cable Length 1 8m 72 uncoiled Display 45 mm 1 8 in digit height LCD display with 4 color backlight Power cord length 1 9m 75 fully extended Scale Base Size 316mm L x 280mm W x 45mm H 12 4 L x 11 W x 1 77 H Indicator Size 252 x 176...

Страница 11: ......

Страница 12: ... the QR code or by going to the product page and from there download it Go to www ohaus com to register your product s warranty Product Page http www ohaus com Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 1 973 377 9000 Fax 1 973 944 7177 With offices worldwide www ohaus com 30728940 P N 30728940A 2022 Ohaus Corporation all rights reserved QR Code ...