QIAcube Connect MDx User Manual 12/2020
138
7a.
To cut the O-Ring, open the hood and rotate the grey lever counter-clockwise until you
reach the black stop.
Open the hood
Rotate the grey lever counter-clockwise
7b.
Open the grey lever and remove the O-Ring (by using the tweezers) from pipetting
channel.
Note
: If required, repeat cutting process until O-Ring is cut completely and can be
removed.
Open the grey lever and remove the O-Ring using the tweezers