QIAcube Connect MDx User Manual 12/2020
142
9
Appendix A – Technical data
QIAGEN reserves the right to change specifications at any time.
9.1
Operating conditions
Power
100–240 V AC, 50/60 Hz, 650 VA Mains supply voltage
fluctuations are not to exceed 10% of nominal supply voltages.
Note
: The apparent power can exceed 650 VA for up to 2 seconds
during the centrifuge acceleration and can reach an approximate
value of 1000 VA.
Fuse
2x T8A L 250V
Overvoltage
category
II
Air temperature
20 to 25ºC (68 to 77ºF)
Relative humidity
20–75% (noncondensing)
Altitude
Up to 2000 m (6500 ft.)
Place of operation
For indoor use only
Pollution level
2
Environmental class
3K20 (IEC 60721-3-3)
9.2
Transport conditions
Air temperature
–25ºC to 60ºC (–13ºF to 140ºF) in manufacturer’s package
Relative humidity
Max. 75% (noncondensing)
9.3
Storage conditions
Air temperature
5ºC to 40ºC (41ºF to 104ºF) in enclosed location
Relative humidity
Max. 75% (noncondensing)