QIAcube Connect MDx User Manual 12/2020
127
Comments and suggestions
Robotic arm
Robotic arm does not return to set
position
Ensure that the instrument is placed on a stable, flat and
level surface as described in section 4.1.1, Site
requirements. In other cases, contact QIAGEN Technical
Services.
Instrument loading
In case of incorrect instrument loading, thoroughly read
the error message. It will guide you to the
missing/wrong item.
Pipettor
Pipet tips not picked up by
automatic pipettor
Make sure that the tip rack is not damaged and is
correctly positioned on the worktable.
Pipet tips not disposed of correctly
Empty the waste drawer and ensure it is not broken.
Check that tip disposal slot is not damaged or
obstructed. Perform regular maintenance, as described
in section 6.3
Droplets observed on worktable
The pipettor is dripping liquid. Check that the reagent
bottles contain the correct buffers and are correctly
placed in the reagent bottle rack. Be sure to use the right
plasticware. Check the volumes in the sample tubes and
tubes of accessory buffer(s), if applicable. Do not exceed
the recommended amount of starting material to avoid
blocking disposable filter-tips. If tip racks have been
refilled, make sure that the correct tips were used.
Check tightness of pipettor in maintenance section as
described in section 6.7.2, Tightness test. If leakage
could be detected, change O-Ring as described in
section 7.3.5, O-ring exchange. If issue persists, contact
QIAGEN Technical Services.
Mechanical
Frame of instrument is distorted
(e.g., uneven, unstable or not level)
Ensure that the instrument is placed on a stable, flat and
level surface as described in section 4.1.1, Site
requirements.
Hood sensor error: instrument will
not function
Ensure the hood is properly closed. The instrument will
not function if the hood is open.
Broken instrument hood
Ensure that only the cleaning products as described in
section 6, Cleaning and Maintenance are used on the
hood.