QIAcube Connect MDx User Manual 12/2020
111
6.3
Regular maintenance
After running a protocol, perform the regular maintenance procedure described below.
Run completed screen
1.
Open the waste drawer and empty tips and columns (if necessary) into a suitable laboratory
waste container.
2.
Remove used disposable labware and unwanted samples and reagents from the worktable.
Discard them according to your local safety regulations.
Note
: If the robotic arm prevents you from reaching a position, do not move the robotic arm
manually. Instead, proceed as follows:
Press
Move left
or
Move right
on the Run Completed screen, as needed. The robotic arm will
start to move. The hood can remain open during this movement.
Ensure that you stand clear of the instrument while the robotic arm is moving. Wait until the
robotic arm has completed its movements.
3.
Replace the lids of the reagent bottles and close tightly. Store the bottles according to the
instructions in the relevant kit handbook.
You can now run another protocol or switch off the QIAcube Connect MDx.
6.4
Daily maintenance
After running the last protocol of the day, perform the daily maintenance procedure. The software
guides you through each step to be performed: