QIAcube Connect MDx User Manual 12/2020
18
WARNING
Risk of explosion
[W23]
When cleaning the QIAcube Connect MDx with alcohol-based disinfectant,
leave the QIAcube Connect MDx hood open to allow flammable vapors to
disperse.
Only clean the QIAcube Connect MDx when worktable components have
cooled down.
WARNING
Risk of fire
[W24]
Do not allow cleaning fluid or decontamination agents to come into contact
with the electrical parts of the QIAcube Connect MDx.
WARNING
Risk of personal injury and material damage
[W25]
To prevent the rotor nuts from loosening during operation of the centrifuge,
securely tightened the nuts using the rotor key supplied with QIAcube Connect
MDx.
WARNING
UV irradiation hazard
[W26]
A mechanical lock ensures that the hood must be closed for operation of the
UV LED.
If the hood sensor or lock is not functioning correctly, contact QIAGEN
Technical Services.
WARNING
Risk of personal injury and material damage
[W27]
Make sure that lids from spin columns and 1.5 ml microcentrifuge tubes are in
the correct position and pushed all the way down to the bottom of the slots on
the sides of the rotor adapter. Incorrectly positioned lids can break off during
centrifugation.
WARNING
Risk of personal injury and material damage
[W28]
Be sure the lid is completely removed from the spin column. Spin columns with
partially removed lids may not be removed properly from the rotor, causing
the protocol run to crash.
CAUTION
Damage to the instrument
[C10]
Do not use bleach, solvents, or reagents containing acids, alkalis or abrasives
to clean QIAcube Connect MDx.