QIAcube Connect MDx User Manual 12/2020
26
USB ports
The QIAcube Connect MDx has four USB ports. Two are located on the left of the touchscreen and
two are located behind the touchscreen.
The USB ports located on the left of the touchscreen allow connection of the QIAcube Connect MDx
to a USB flash drive. Data files, such as support package, protocols or report files, can be
transferred via the USB port from the QIAcube Connect MDx to the USB flash drive. The USB ports
can also be used to plug in the supplied external barcode scanner.
USB ports located below the touchscreen, allow insertion of a Wi-Fi adapter to enable Wi-Fi
connection to a local network.
Important
: Use only the USB flash drive provided by QIAGEN. Do not connect other USB flash
drive devices to USB ports.
Important
: Do not remove the USB flash drive while downloading or transferring data or software
to or from the instrument.
Important
: Always turn off the QIAcube Connect MDx in order to plug or unplug the
Wi-Fi USB device. Plug-and-play of the Wi-Fi USB device while the instrument is turned on is
not supported.
Waste drawer
Used disposable filter-tips are discarded through two slots in the worktable and collected in the
waste drawer. Used columns (e.g., QIAshredder columns) are also discarded into this drawer.
CAUTION
Damage to the instrument
[C3]
Empty the tip disposal container prior to use to prevent a tip jam in the waste
drawer. Failure to empty the waste container may block the robotic arm that
could cause run failure or instrument damage.
WARNING
Hazardous chemicals and infectious agents
[W16]
The waste may contain toxic material and must be disposed of properly. Refer
to your local safety regulations for proper disposal procedures.