QIAcube Connect MDx User Manual 12/2020
139
8.
Close the hood and press
Next
to pick up the prepared new O-Ring.
9.
Open the hood and visually check if new O-Ring sits firmly on tip adapter.
Check if the new O-Ring sits firmly
Note
: If the O-Ring was not successfully picked up, complete the O-Ring replacement
procedure and restart.
10.
Close the hood press
Next
to remove the O-Ring change tool.
11.
Open the hood and remove the O-Ring change tool.
12.
Wipe and clean the O-Ring change tool with alcohol-based disinfection wipes. Incubate as
appropriate, rinse thoroughly with distilled water and wipe dry with lint-free paper towels.
13.
Press
Done
to complete the O-Ring replacement. The date of the last performed O-Ring
replacement is updated automatically.
WARNING
Risk of personal injury and material damage
[W1]
Improper use of the QIAcube Connect MDx may cause personal injuries or
damage to the instrument. The QIAcube Connect MDx must only be
operated by qualified personnel who have been appropriately trained.
Servicing of the QIAcube Connect MDx must only be performed by a
QIAGEN Field Service specialist.